Remote Control Commands
R&S
®
NRX
213
User Manual 1178.5566.02 ─ 05
Parameters:
<port>
INTernal | EXT2 | EXTernal2
*RST:
INTernal
Manual operation:
See
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:STATe]
<state>
Enables or disables the trigger master mode of the power sensor. In this state, the
power sensor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer:PORT
.
Typically, the trigger master uses its internal trigger source. But you can also trigger the
trigger master externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger master externally, use
EXTernal1
as external trigger
input port (trigger source) and
EXTernal2
as trigger master output port or vice versa.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger<undef>:ALL:SLOPe
<slope>
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
<slope>
Effective only if
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
INTernal
or
EXTernal
.
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Measurement Settings and Results