Remote Control Commands
R&S
®
NRX
208
User Manual 1178.5566.02 ─ 05
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<count>
Range:
Depends on power sensor
*RST:
1
TRIGger<undef>:ALL:DELay:AUTO
<stat>
TRIGger<Measurement>[:CHANnel<Channel>]:DELay:AUTO
<stat>
If enabled, no measurement is started until the power sensor has settled. For this pur-
pose, the delay value is automatically determined.
If a longer period is set using
TRIGger<Measurement>[:CHANnel<Channel>]:
, the automatically determined delay is ignored.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<stat>
ON | OFF
*RST:
OFF
TRIGger<undef>:ALL:DELay[:VALue]
<delay>
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
<delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Measurement Settings and Results