Remote Commands
R&S
®
FSW
1168
User Manual 1173.9411.02 ─ 43
Manual operation:
See
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
This command moves a delta marker to the next higher minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
14.8.3.5
Retrieving Marker Results
The following commands are used to retrieve the results of markers.
You can use the marker values to position the center frequency or reference level
directly using the following commands:
●
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
●
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
Useful commands for retrieving results described elsewhere:
●
●
●
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt?
●
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?
●
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
●
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
●
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
●
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
●
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
●
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?
●
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
●
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:QFACtor?
●
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
Analyzing Measurements (Basics)