![R&S FSW Series User Manual Download Page 1196](http://html2.mh-extra.com/html/rands/fsw-series/fsw-series_user-manual_31913991196.webp)
Remote Commands
R&S
®
FSW
1196
User Manual 1173.9411.02 ─ 43
.............................................................................. 1196
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
.......................................................... 1196
CALCulate<n>:MARKer<m>:COUNt:RESolution
............................................................. 1197
CALCulate<n>:MARKer<m>:COUNt
<State>
This command turns the frequency counter at the marker position on and off.
The frequency counter works for one marker only. If you perform a frequency count
with another marker, the R&S
FSW deactivates the frequency count of the first marker.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation:
See
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Before you can use the command, you have to turn on the frequency counter.
Analyzing Measurements (Basics)