Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
199
User Manual 1173.9240.02 ─ 23
Conversion Loss Table Settings
The following settings are required to configure and manage conversion loss tables.
....................................................................................199
..................................................................................... 199
[SENSe:]CORRection:CVL:CATAlog?
..............................................................................200
................................................................................... 200
[SENSe:]CORRection:CVL:COMMent
..............................................................................200
.................................................................................... 201
[SENSe:]CORRection:CVL:HARMonic
............................................................................. 201
.................................................................................... 201
.................................................................................. 202
.................................................................................. 202
[SENSe:]CORRection:CVL:SNUMber
.............................................................................. 202
[SENSe:]CORRection:CVL:BAND
<Type>
This command defines the waveguide band for which the conversion loss table is to be
used. This setting is checked against the current mixer setting before the table can be
assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Band>
K | KA | Q | U | V | E | W | F | D | G | Y | J | USER
Standard waveguide band or user-defined band.
For a definition of the frequency range for the pre-defined bands,
see
*RST:
F (90 GHz - 140 GHz)
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BAND KA
Sets the band to KA (26.5 GHz - 40 GHz).
Manual operation:
See
[SENSe:]CORRection:CVL:BIAS
<BiasSetting>
This command defines the bias setting to be used with the conversion loss table.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<BiasSetting>
numeric value
*RST:
0.0 A
Default unit: A
Configuring the Measurement