Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
200
User Manual 1173.9240.02 ─ 23
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BIAS 3A
Manual operation:
See
See
[SENSe:]CORRection:CVL:CATAlog?
This command queries all available conversion loss tables saved in the
C:\r_s\instr\user\cvl\
directory on the instrument.
This command is only available with option B21 (External Mixer) installed.
Usage:
Query only
[SENSe:]CORRection:CVL:CLEAr
This command deletes the selected conversion loss table. Before this command can
be performed, the conversion loss table must be selected (see
This command is only available with option B21 (External Mixer) installed.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:CLE
Usage:
Event
Manual operation:
See
[SENSe:]CORRection:CVL:COMMent
<Text>
This command defines a comment for the conversion loss table. Before this command
can be performed, the conversion loss table must be selected (see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Text>
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:COMM 'Conversion loss table for
FS_Z60'
Manual operation:
See
Configuring the Measurement