Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
192
User Manual 1173.9240.02 ─ 23
Manual operation:
See
See
See
11.4.2.2
Using External Mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the R&S
FSW-B21 option to be
installed and an external mixer to be connected to the front panel of the R&S
FSW.
In MSRA / MSRT mode, external mixers are not supported.
For details on working with external mixers see the R&S
FSW User Manual.
Conversion Loss Table Settings
........................................................................... 199
Programming Example: Working with an External Mixer
......................................203
Basic Settings
The basic settings concern general usage of an external mixer.
............................................................................................. 192
............................................................................................193
................................................................................................193
............................................................................................. 194
[SENSe:]MIXer[:STATe]
<State>
Activates or deactivates the use of a connected external mixer as input for the mea-
surement. This command is only available if the optional External Mixer is installed and
an external mixer is connected.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
MIX ON
Manual operation:
See
[SENSe:]MIXer:BIAS:HIGH
<BiasSetting>
This command defines the bias current for the high (second) range.
This command is only available if the external mixer is active (see
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
Configuring the Measurement