Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
193
User Manual 1173.9240.02 ─ 23
Manual operation:
See
[SENSe:]MIXer:BIAS[:LOW]
<BiasSetting>
This command defines the bias current for the low (first) range.
This command is only available if the external mixer is active (see
Parameters:
<BiasSetting>
*RST:
0.0 A
Default unit: A
Manual operation:
See
[SENSe:]MIXer:LOPower
<Level>
This command specifies the LO level of the external mixer's LO port.
Parameters:
<Level>
numeric value
Range:
13.0 dBm to 17.0 dBm
Increment: 0.1 dB
*RST:
15.5 dBm
Example:
MIX:LOP 16.0dBm
Manual operation:
See
[SENSe:]MIXer:SIGNal
<State>
This command specifies whether automatic signal detection is active or not.
Note that automatic signal identification is only available for measurements that per-
form frequency sweeps (not in vector signal analysis or the I/Q Analyzer, for instance).
Parameters:
<State>
OFF | ON | AUTO | ALL
OFF
No automatic signal detection is active.
ON
Automatic signal detection (Signal ID) is active.
AUTO
Automatic signal detection (Auto ID) is active.
ALL
Both automatic signal detection functions (Signal ID+Auto ID)
are active.
*RST:
OFF
Manual operation:
See
See
Configuring the Measurement