Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
17
Operating Manual 1176.7561.02 ─ 06
Display Mode
●
●
●
●
●
●
For details see
Chapter 2.1.7, "Trace Mode Overview"
Detector Auto Select
Activates automatic detector selection (see
activated, the "Trace Detector" setting is ignored.
Trace Detector
Defines a specific trace detector. If one of the following settings is
defined, the "Detector Auto Select" option is deactivated.
●
●
●
●
●
●
●
●
3. To configure several traces to predefined display modes in one step, press the but-
ton for the required function:
Preset All Traces
Trace 2-6:
Select Max | Avg | Min
Trace 4-6:
Select Max | ClrWrite | Min
Trace 4-6:
Chapter 2.1.7, "Trace Mode Overview"
2.1.7
Trace Mode Overview
The traces can be activated individually for a measurement or frozen after completion
of a measurement. Traces that are not activate are hidden. Each time the trace mode
is changed, the selected trace memory is cleared.
The R&S
FSVA/FSV offers 6 different trace modes:
Clear Write
Overwrite mode: the trace is overwritten by each sweep. This is the default setting.
All available detectors can be selected.
Remote command:
DISP:TRAC:MODE WRIT
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Instrument Functions Analog Demodulation