This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE AVER
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
View ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
The current contents of the trace memory are frozen and displayed.
Note:
If a trace is frozen, the instrument settings, apart from level range and reference
level (see below), can be changed without impact on the displayed trace. The fact that
the displayed trace no longer matches the current instrument setting is indicated by the
icon on the tab label.
If the level range or reference level is changed, the R&S
FSVA/FSV automatically
adapts the measured data to the changed display range. This allows an amplitude
zoom to be made after the measurement in order to show details of the trace.
Remote command:
DISP:TRAC:MODE VIEW
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Blank ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Hides the selected trace.
Remote command:
DISP:TRAC OFF
, see
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Opens a submenu to select the detector manually, or activate automatic selection.
If a detector was selected manually, the "MAN" indicator is highlighted.
If "AUTO" is selected, the detector is defined automatically, depending on the selected
trace mode:
Trace mode
Detector
Clear Write
Auto Peak
Max Hold
Positive Peak
Min Hold
Negative Peak
Average
Sample Peak
View
–
Blank
–
For details see
Chapter 2.1.8, "Detector Overview"
Note:
In Analog Demod mode, if AUTO is selected, the Auto Peak detector is used
regardless of the trace mode. However, if Noise or Phase Noise measurements are
performed in Analog Demod mode, the Sample Detector is used.
Auto Select ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the best detector for the selected trace and filter mode. This is the default set-
ting.
Softkeys of the Analog Demodulation option