Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
19
Operating Manual 1176.7561.02 ─ 06
Remote command:
DISP:TRAC:MODE VIEW
, see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Blank
Hides the selected trace.
Remote command:
DISP:TRAC OFF
, see
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
2.1.8
Detector Overview
The measurement detector for the individual display modes can be selected directly by
the user or set automatically by the R&S
FSVA/FSV. The detector activated for the
specific trace is indicated in the corresponding trace display field by an abbreviation.
The detectors of the R&S
FSVA/FSV are implemented as pure digital devices. They
collect signal power data within each measured point during a sweep. The default
number of sweep points is 691. The following detectors are available:
Table 2-3: Detector types
Detector
Indicator
Function
Auto Peak
Ap
Determines the maximum and the minimum value
within a measurement point (not available for SEM)
Positive Peak
Pk
Determines the maximum value within a measure-
ment point
Negative Peak (min peak)
Mi
Determines the minimum value within a measure-
ment point
RMS
Rm
Determines the root mean square power within a
measurement point
Average
Av
Determines the linear average power within a mea-
surement point
Sample
Sa
Selects the last value within a measurement point
The result obtained from the selected detector within a measurement point is displayed
as the power value at this measurement point.
All detectors work in parallel in the background, which means that the measurement
speed is independent of the detector combination used for different traces.
Instrument Functions Analog Demodulation