General data
Display
14 cm (5.7“) LC color display
Resolution
320 × 240 pixel
Memory
Settings and traces
CMOS RAM
up to 256
Environmental conditions
Temperature
Operating temperature range
¸FSH powered from internal battery
¸FSH powered from AC power supply
0 °C to +50 °C
0 °C to +40 °C
Storage temperature range
–20 °C to +60 °C
Battery charging mode
0 °C to +40 °C
Climatic conditions
Relative humidity
95 % at +40 °C (EN 60068)
IP class of protection
51
Mechanical resistance
Vibration, sinusoidal
meets EN 60068-2-1, EN 61010-1
5 Hz to 55 Hz: max 2 g, 55 Hz to 150 Hz: 0.5 g constant,
12 minutes per axis
Vibration, random
meets EN 60068-2-64, 10 Hz to 500 Hz, 1.9 g,
30 minutes per axis
Shock
meets EN 60068-2-27, 40 g shock spectrum
RFI suppression
meets EMC directive of EU (89/336/EEC)
and German EMC legislation
Immunity to radiated interference
Level display at 10 V/m (reference level
≤
–10 dBm)
Input frequency
IF
Other frequencies
10 V/m
<–75 dBm (nominal)
<–85 dBm (nominal)
< displayed noise level
Power supply
AC supply
plug-in AC power supply (¸FSH-Z33)
100 V AC to 240 V AC, 50 Hz to 60 Hz, 400 mA
External DC voltage
15 V to 20 V
Internal battery
NiMH battery, type Fluke BP190 (¸FSH-Z32)
Battery voltage
6 V to 9 V
Operating time with fully-charged battery
4 h with tracking generator off,
3 h with tracking generator on
Lifetime
300 to 500 charging cycles
Power consumption
typ. 7 W
Safety
meets EN 61010-1, UL 3111-1, CSA C22.2 No. 1010-1
Test mark
VDE, GS, CSA, CSA-NRTL
Dimensions (W × H × D)
170 mm × 120 mm × 270 mm
Weight
2.5 kg
21
Handheld Spectrum Analyzer ¸FSH