Running an EMI Band Evaluation Test
01.00
Rohde & Schwarz
EMI Automotive Band Evaluation with R&S©EMC32-K51
– 1SP07-
32
Go to Lowest
Frequency
By clicking this button the frequency will jump to the Start frequency, which is
the lowest most frequency defined in the Test Template e.g. if Start
Frequency equals 30MHz then by pressing this button it will jump to
start/lowest frequency 30MHz.
Switch to Automatic
Scan / Sweep
By clicking this button it will switch to automatic Scan / Sweep measurement.
Switching to this mode the whole range of the frequency shall be scanned or
swept as defined in the test template.
Switch to Single
Measurement
By pressing this button the mode will be switched to the single measurement
mode, where one measurement will be done for each single frequency.
Switching to the single measurement in only possible while the test is
stopped.
Calculate Limit and
Margin
By clicking this button, a dialog window will be opened. This dialog window
provides all necessary options to calculate limit and margin values for a result
table which will be available after the measurement.
The option which are available in the dialog window are:
1. Result type
o
Spectrum:this is the resultant spectrum produced after the test.
o
Single Measurement results: the result of a single measurement.
o
From this Test: This dialog will select the table from the current
test.
2. Evaluation Column
o
A
detector trace
column, from the resultant table (single or
scan/sweep results, evaluation results), can be selected here.
3. Limit Lines
o
From this Test: Use limit line column from the limit lines table of
current measurement.
o
Global: Limit lines defined from the Global Standards.
4. Output File:
o
A name can be give to the output file which will be generated after