Instrument tour
R&S
®
EDS300
33
Getting Started 1179.5630.02 ─ 01
5.2.5
USB
The rear panel provides additional female USB (USB-A) connectors to connect a
memory device or power sensor.
The connectors are standard USB 3.0 ports with a maximum transmission rate of
40
Mbit/s.
For further USB connectors, see also
Chapter 5.1.4.2, "USB interface"
5.2.6
REF 10 MHz IN/OUT
The "REF 10 MHz IN/OUT" connector can provide an external reference signal to
the R&S
EDS300, or from the R&S
EDS300 to a connected device. In either
case, the reference signal is at 10
MHz, with a power level of 1
V
eff
. The connec-
tor is a 50
Ω
BNC socket.
Configure whether the connector is used for input or output in the general instru-
ment settings.
5.2.7
LAN interface
Use the "LAN" interface to connect the R&S
EDS300 to a local network for
remote control, printouts or data transfer. The data can be transferred with a rate
of up to 1
Gbit per second. Configure the IP address and subnet mask in the gen-
eral instrument settings.
5.2.8
DVI
You can connect an external (TFT) monitor or other display device to the
R&S
EDS300's DVI (Digital visual interface) to provide an enlarged display. The
interface consists of a DVI-D socket (24+1).
Chapter 4.8.4, "Connecting an external monitor"
5.2.9
RS232 GPS interface
2-port RS232, 9-pin D-Sub connector for a GPS receiver providing NMEA proto-
col data. The NMEA protocol data is displayed and stored with the recorded data.
Rear panel view