Remote control commands
R&S
®
AREG800A
357
User Manual 1179.3615.02 ─ 01
Example:
:SENS1:POW:OFFS 0.4dB
Sets a level offset of 0.4 dB
:SENS1:POW:OFFS:STAT ON
A level offset of 0.4 dB is added to the measured value.
Manual operation:
See
"Level Offset State,Level Offset"
:SENSe<ch>[:POWer]:SNUMber?
Queries the serial number of the sensor.
Return values:
<SNumber>
string
Example:
:SENS1:SNUM?
Queries the serial number.
Usage:
Query only
Manual operation:
See
"Sensor type and serial number"
:SENSe<ch>[:POWer]:SOURce
<Source>
Determines the signal to be measured.
Note:
When measuring the RF signal, the sensor considers the corresponding correc-
tion factor at that frequency, and uses the level setting of the instrument as reference
level.
Parameters:
<Source>
A | USER | RF
*RST:
A
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:STATus[:DEVice]?
Queries if a sensor is connected to the instrument.
Return values:
<Status>
0 | 1 | OFF | ON
*RST:
0
Example:
:SENS1:STAT?
Response: 1
A sensor is connected.
Usage:
Query only
Manual operation:
See
SENSe, READ, INITiate and SLISt subsystems