SOURce:VOLTage[:STATe]
<State>
This command turns DC power sources on and off in general.
When you turn off the DC power sources, no power is supplied even when you have
turned on one of the sources individually with one of the following commands.
●
SOURce:VOLTage:AUX:LEVel[:STATe]
●
SOURce:VOLTage:CONTrol<i>:LEVel[:STATe]
●
SOURce:VOLTage:POWer<i>:LEVel[:STATe]
Note that you can turn on the global power supply if at least one of the individual sup-
plies has been turned on.
Parameters:
<State>
ON | 1
DC power sources are ready for use.
OFF | 0
DC power sources are turned off.
*RST:
OFF
Example:
//Turn on signal source
SOUR:VOLT:AUX:LEV ON
SOUR:VOLT ON
Manual operation:
See
"Turning the DC output on and off"
11.6.3
Remote Commands to Configure User Ports
.................................................................................................348
...............................................................................................349
............................................................................................349
..........................................................................................349
INPut<ip>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSPN.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST:
1
Common Measurement Settings