Remote Control
R&S
®
FSPN
349
User Manual 1179.4363.02 ─ 01
INPut<ip>:UPORt[:VALue]
This command queries the control lines of the user ports.
For details see
Suffix:
<ip>
.
1 | 2
irrelevant
Return values:
<Level>
bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range:
#B00000000 to #B00111111
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.
OUTPut<up>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSPN.
Suffix:
<up>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
User port is switched to INPut
ON | 1
User port is switched to OUTPut
Example:
OUTP:UPOR:STAT ON
OUTPut<up>:UPORt[:VALue]
<Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit
7
6
5
4
3
2
1
0
Pin
N/A
N/A
5
3
4
7
6
2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Suffix:
<up>
.
irrelevant
Common Measurement Settings