Rainin Pipetman P-10 Manual Download Page 8

6

Pre-Rinsing Recommended

Some solutions (e.g. serum, protein-containing solutions,
and organic solvents) can leave a film on the inside tip
wall, resulting in an error larger than the tolerance speci-
fied. Since this film remains relatively constant in succes-
sive pipettings with the same tip, excellent precision may
be obtained by refilling the tip and using the refilled vol-
ume as the sample. Successive samples from this same
tip will exhibit good reproducibility (low variance).

Reverse Mode Pipetting

Another way of reducing error due to film retention is re-
verse mode pipetting; the operating sequence is reversed:

1. Mount a disposable tip on the pipette shaft.

2. Press the pushbutton fully to the second stop.

3. Immerse the tip in liquid and return the plunger slowly

to the full up position. Wait a moment for the liquid
column to reach equilibrium in the tip.

4. Wipe any excess liquid from the outside of the tip

without touching the orifice.

5. To dispense, rest the end of the tip against the vessel

wall and press the plunger to the first stop. Hold this
position a few seconds, or long enough for the liquid
column to reach equilibrium again.

6. Remove the tip from the receiving vessel without

blowing out the remaining liquid.

7. Return excess sample in the tip to the original sample

container, if desired. Discard the used tip.

Pipetting Liquids of Varying Density

Pipetman lets you compensate for solutions of density
much different from water, by setting the volume slightly
higher or lower than that required. The compensation
amount must be determined empirically.

E.g., if pipetting 10 µL of CsCl solution, you determine
that the volume delivered is actually 8.5 µL (

5 samples).

Try changing the volume setting to 11.8 µL and repeat the
measurements. If the volumes delivered are still not close
enough to 10 µL, make another slight volume adjustment
until the measurements are as desired.

Very dense liquids may not be suitable for air displace-
ment pipetting. Use Microman® positive displacement
pipettes for these liquids.

pipetman rev I  9/13/00  3:12 PM  Page 8

Summary of Contents for Pipetman P-10

Page 1: ...Pipetman continuously adjustable digital microliter pipettes 8 pipettes for volume ranges from 0 1 L to 10 mL P 2 P 10 P 20 P 100 P 200 P 1000 P 5000 P 10ML Pipetman P 200 shown...

Page 2: ...ons 11 Replacement Parts Back Cover FIGURES Figure 1 Pipetman P 200 1 Figure 2 Operating Pipetman 4 Figure 3 Removing Tip Ejector Arm 9 DIAGRAMS FOR REPLACEMENT PARTS Models P 2 P 10 P 20 P 100 12 Mod...

Page 3: ...s in microliters The digital volume indicator simplifies volume setting and virtually eliminates calculation errors All Pipetman models have a highly polished stainless steel piston patented polyethyl...

Page 4: ...are shown below Pipetman Range L Smallest Model Adjustable Recommended Increment L P 2 0 to 2 0 1 to 2 0 002 P 10 0 to 10 0 5 to 10 0 02 P 20 0 to 20 2 to 20 0 02 P 100 0 to 100 10 to 100 0 2 P 200 0...

Page 5: ...e The friction ring prevents unintentional volume changes 3 Attach a new disposable tip to the pipette shaft Press only enough to make a positive airtight seal 4 Press the plunger to the FIRST STOP Th...

Page 6: ...d flexible so that the shaft is not scratched or worn prematurely Tips must be free from microscopic flash and particu lates Flash gives poor precision and accuracy The tip orifice must be the correct...

Page 7: ...is important If exceeded the vol ume measured will be inaccurate possibly out of specifi cation Tip angle is also important Hold the pipette verti cally or within 20 degrees of vertical Pipetting Gui...

Page 8: ...he tip without touching the orifice 5 To dispense rest the end of the tip against the vessel wall and press the plunger to the first stop Hold this position a few seconds or long enough for the liquid...

Page 9: ...contact with corrosive fumes may corrode the piston This will result in premature seal wear and may require refinishing or replacement of the piston Exposure of internal Pipetman components to corrosi...

Page 10: ...05803D which use a grease seal Leaks Inaccurate Sampling 1 Loosened shaft Tighten coupling by hand 2 Split or cracked shaft Remove the tip ejector and inspect the shaft for fracture or split end Repla...

Page 11: ...jector opening over the end of the Pipetman shaft 5 Push the plastic collar into the handle until the tip ejector is securely in place Make sure that tips will fit properly on the shaft Figure 3 Tip E...

Page 12: ...U S only from Rainin It is NOT necessary to recalibrate the pipette after changing the seal or shaft Recalibration of the pipette is only necessary when the piston is replaced and should only be done...

Page 13: ...ipetman are defined in this manual as follows Accuracy is the closeness of a measured volume to the true volume as specified by the volume setting of the pipette Also known as mean error Precision is...

Page 14: ...e diagrams are for dual adjustment Pipetman and cannot be used on older Pipetman models For convenience old style part numbers are also listed on the back page 12 P 2 P 10 P 20 P 100 A B M L K J H G F...

Page 15: ...13 P 200 P 1000 P 5000 P 10ML A B M L K J H G F C D E A B G F C D E A B D N N A B F G D G F...

Page 16: ...00 n a n a n a n a n a n a 6190 164 6190 164 Filters 1000 n a n a n a n a n a n a 6190 165 6190 165 This column is for P 10ML after S N N05806D For P 10ML pre S N N05803D same as listed above except P...

Reviews: