SECTION 3 OPERATION OF ACCESSORIES
3 - 11
3.7 Dilution Probe
A dilution probe device is offered as an accessory to the
MiniRAE Plus Professional PID unit. This device mixes one
part of incoming gas sample with 9 parts of clean air to dilute
the concentration of the gas sample by 10 times. The clean air
is obtained through a charcoal filter.
Install dilution probe :
Remove the regular probe. Make sure that O-ring and a
membrane filter remain inside the hole on the front of the sensor
block. Screw on the dilution probe.
Calibrate dilution probe :
In the factory, the dilution probe is preset at approximately 10:
1 ratio with about 5-10% error. User can adjust the dilution
probe during calibration using a 5/64" Allen wrench to achieve
exact 10:1 ratio. Here is how:
Step 1 : Connect the "zero gas" to the probe, follow the
instructions in section 4.8.1 to calibrate the zero gas.
Step 2 : Follow the instruction in section 4.8.2 to enter the
standard reference gas value.
Note: enter only 1/10 of the
actual value of the standard gas, not the actual value.
For
example, if the calibration gas is 500 ppm, enter 50 in this step,
do not enter 500.
Step 3 : Connect the standard reference gas to the probe.
Turn on the gas, press [enter] key on the MiniRAE. Wait for
about 30 seconds until the gas reading appear on the LCD
Summary of Contents for MiniRAE Plus PGM-76
Page 2: ...RAE Systems Inc 680 West Maude Ave 1 Tel 408 481 4999 Sunnyvale CA 94086 Fax 408 481 4998...
Page 14: ......
Page 18: ...SECTION 1 GENERAL INFORMATION 1 4 This page is intentionally left blank...
Page 56: ...SECTION 3 OPERATION OF ACCESSORIES 3 16 This page is intentionally left blank...
Page 104: ...SECTION 5 PROGRAMMING IN SURVEY MODE 5 18 This page is intentionally left blank...