SECTION 2 OPERATION OF MINIRAE PLUS
2 - 14
2.7 Preset Alarm Limits and
Calibration
The MiniRAE Plus Professional PID Unit is factory calibrated
with standard isobutylene gas, and is programmed with default
alarm limits as listed below. Refer to section 4 and 5 for
programming procedures if new calibration or new alarm limits
are required.
Table 2.2
Factory Calibration and Preset Alarm Limits
Calibration Standard Gas
100 ppm isobutylene
Hygiene Mode:
TWA Alarm Limit
10 ppm
STEL Alarm Limit
25 ppm
Peak Alarm Limit
50 ppm
Survey Mode:
Low Alarm Limit
10 ppm
High
Alarm Limit
100 ppm
Battery Low Level
5.5 volt
Power Off Level
5.3 volt