RESPONSE INFORMATION
Calibration factors for a number of commonly used
organic vapor for 10.6 eV lamp is included in Table
C.1
Table C.1
Correction Factors for the 10.6 eV lamp:
Compound
Correction
Factor
Acetaldehyde
5.5
Acetone
1.1
Allyl alcohol
2.4
Benzene
0.53
Butadiene
1.0
Butyl acetate
2.6
Carbon disulfide
1.2
Cyclohexane
1.4
Cyclohexanone
1.0
Cyclohexene
0.90
Diethyl amine
0.97
Diethyl sulfide
0.51
Ethyl acetate
4.6
Ethyl alcohol
12
Ethyl benzene
0.52
Ethyl ether
1.1
Ethylene
9.9
Heptane (n-)
2.6
Hexane
4.2