PROGRAMMING
4-15
4.4.6
Change Correction Factor
This function allows the user to change the Correction
Factor of the standard calibration gas (only for Cal
Memory #0).
1. “Change Correction Factor?” is the sixth sub-menu in
the Calibration sub-menu.
2. Press
[Y/+]
key. Display shows the gas name, then the
correction factor.
A cursor blinks at the left-most digit of the correction
factor. If user wants to modify the correction factor, go
to Step 3. Otherwise, press and hold the
[MODE]
key
for 1 second to accept the previously stored correction
factor value and return to the first sub-menu of the
calibrate/select gas menu.
3.
Starting from the left-most digit of the correction
factor, use
[Y/+]
or
[N/-]
key to change the digit value
and press
[MODE]
key momentarily to advance to the
next digit, the cursor will move to the next digit to the
right. Repeat this process until all digits are entered.
Press and hold the
[MODE]
for 1 second to exit.
4.
The display shows “Save?” To confirm the new value,
press
[Y/+]
to accept the change. Press
[N/-]
or
[MODE]
to discard the change and return to the first
sub-menu, Calibrate and Select Gas.