24
10. PERIPHERAL DEVICES
10.1. IM01.EX Communication Module
IM01.EX communication module is a device intended for operation outside
hazardous area, it is equipped with intrinsically safe circuits which can be
placed in zones endangered with explosion. The module can be connected, via
the intrinsically safe interface, with HX5.EX scale placed in the hazardous area.
This allows integration of HX5. EX scale with various accessories: barcode
scanners, printers, control buttons, light signalling towers, buzzers, PLC
controllers and other controlling/signalling devices, PCs operating outside
hazardous areas.
10.1.1. Connecting IM01.EX to the Mains
IM01.EX communication module is equipped with cable terminated with a plug
featuring a ground pin, plug type is conditioned by region/country. Connect the
cable to the wall outlet with a ground pin.
It is not allowed to connect the IM01.EX communication
module's plug to the wall outlet located in the hazardous
area.
10.1.2. Connecting IM01.EX Module and PUE HX5.EX Indicator
IM01.EX communication module is equipped with a connector for intrinsically
safe circuit, marked as HX5.EX, used to connect PUE HX5.EX indicator.
Connect the PUE HX5.EX indicator to the IM01.EX communication module via
a dedicated cable,
PT0327
.
Connect the IM01.EX communication module to devices
operating in hazardous area when there is no risk of explosive
atmosphere occurrence.
10.1.3. Activation of Connection Between the IM01.EX Module and
HX5.EX Scale
•
Connect the IM01.EX communication module to the HX5.EX scale and to
the mains.
•
Enter
<Peripherals / IM01.EX communication module / Activation>
submenu
and activate the IM01.EX communication module (
- module
active,
- module inactive
).