- 88 -
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
D
H
Space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Give value of max checkweighing threshold
Format:
OUH CR LF
Response:
UH_MASA CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
U
H
Space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Set mass value of a single item (only for PARTS COUNTING)
Format:
SM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
SM_OK CR LF
- command carried out
SM_I CR LF
-
command understood but not
accessible at this moment (e.g. mode
other than PARTS COUNTING)
ES CR LF
- command not recognised (mass
format incorrect)
Set target mass value (e.g. for DOSING)
Format:
TV_XXXXX CR LF
, where:
_ -
space
,
XXXXX
- mass format
Response options:
TV_OK CR LF
- command carried out
TV_I CR LF
-
command understood but not
accessible at this moment (e.g. mode
other than DOSING)
ES CR LF
- command not recognised (mass
format incorrect)
Set reference mass value (e.g. for DEVIATIONS)
Format:
RM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
RM_OK CR LF
- command carried out
RM_I CR LF
-
command understood but not
accessible at this moment (e.g. mode
other than DEVIATIONS)
ES CR LF
- command not recognised (mass
format incorrect)
Value release
Format:
SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command’s function is similar to function of PRINT button located on an overlay,
Internal adjustment
Format: IC CR LF
Response options: