- 71 -
20.4.11. Switch off continuous transmission in current unit
Syntax:
CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
20.4.12. Send all implemented commands
Syntax:
PC CR LF
Possible answers:
PC_- >_Z,T, TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC
– command executed,
the indicator have sent all the implemented commands.
20.5. Manual printouts / automatic printouts
Users can general manual or automatic printouts from the scale.
Manual printouts can be performed after loading the pan
and stabilizing indication by pressing
.
Automatic printouts can be performed only after loading the pan
and stabilizing indication.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1
2
3
4 -12
13
14
15
16
17
18
stability
space
sign
mass
space
unit
CR
LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range