background image

2.  Set  the  controls  of  the  BKF10  as  fol­

lows: 

INPUT  FILTER:  FLAT 

MAX.  INPUT: 

To  suit  output  level  from  device 
under  test.  (10  mV  to  3  V  or  100 mV 
to  30  V.) 

VERNIER:  CAL. 

LEVEL  CONTROL  switch  (16): 

To  suit  output  or  input  level  of  the 
device  under  test. 

LEVEL  AT: 

A)  INPUT  (of  BKF10),  if  a  constant 

output  from  the  device  under 
test  is  required. 

B)  OUTPUT  (of  BKF10),  if a constant 

input  to  the  device  under  test  is 
required. 

Oscillator  function  switch  (8):  MAN. 

OUTPUT  FREQ.  control  (18): 

To  indicate,  say,  1  kHz  on  INPUT 
FREQUENCY  meter. 

3.  Check  that  neither  OUT  OF  RANGE 

lamp  lights. 

4. 

Set  oscillator  function  switch  (8)  to 
SINGLE  SWEEP  and  momentarily  de­

press  the  SINGLE  SWEEP  button  (7). 
This  starts  the  automatic  sweep  func­
tion. 
The  BKF10  sweeps 

downwards 

in  fre­

quency  from  20  kHz  to  20  Hz.  When 
the  sweep  is  completed  (about  60 

seconds)  the  BKF10  will  reset automat­

ically  in  readiness  for  the  next  sweep. 
The  single  sweep  function  may  be 

stopped at  any  time  during  a  sweep  by 

simply  setting  the  oscillator  function 
switch  (8)  to  the  MAN.  position,  which 

may  then  be  reset  to  SINGLE  SWEEP 

by  repeating  item 

(4). 

5.  During  the  sweep  the  three  meters 

show  simultaneously: 

INPUT  FREQUENCY  meter: 

The  frequency  at  which the measure­
ments  are  being  made. 

DISTORTION  meter: 

The  total  harmonic  distortion  at  the 
frequency  indicated  by  the  Input 
Frequency  meter.  (Harmonics  mea­

sured,  10  Hz  to  150  kHz). 

INPUT/OUTPUT  RATIO  meter: 

The  input/output  ratio  of  the  device 

under  test  at  the  frequency  indicat­
ed  by  the  Input  Frequency  meter. 

SIGNAL-TO-NOISE  RATIO 
MEASUREMENTS 

Signal-to-noise measurements are  made  by 
depressing  the  SIN  button  (13)  on  the 

front  panel.  The  bandwidth  of  the  signal­
to-noise  measurements  is  10  Hz  to  150 

kHz.  Depressing  the  SIN  button  holds  the 

ALC  pre-amplifier  gain  constant  while  at 

the  same  time  muting  the  oscillator.  A  dc 
analogue  output  of  the  SIN  measurements 
is  available  as  long  as  the  SIN  button  is 
held  depressed.  Facilities  for  external  con­

trol  of  the  SIN  function  are  available  at 

the  multiconnector  (28).  See  page  5  for 

details. 

INPUT/OUTPUT  RATIO  MEASUREMENTS 

Input/output  ratios  are  indicated  on  the 

linear  dB  scale  of  the  INPUT/OUTPUT 
RATIO  meter. 
The  INPUT/OUTPUT  �I\TIO  meter  is  pro­

vided  with  two  indicator  lamps,  "-dB" 

(4) 

and " +20 dB"  (6),  which  function  as  fol­
lows: 

-dB  Lamp 

(4) 

When  a  negative  dB  value  or  loss  occurs 
in  a  device  under  test,  the  -dB  lamp  will 
light  and  the  value  indicated  on  the  meter 
will  represent  the  loss  directly  in  dB. 

+ 20 dB  Lamp  (6) 

When  the  MAX.  INPUT  switch  is  set  to  the 
30  V  position,  a  20  dB  attenuator  is  con­
nected  into  the  input  circuit  of  the  BKF10. 
This  will  result  in  a  20  dB  lower  reading 

on  the  INPUT/OUTPUT  RATIO  meter  and 
on  the  associated  recorder  output.  As  a 
reminder,  the " + 20 dB"  lamp  lights  when­
ever  the  MAX.  INPUT  switch  is  set  to 

"30  V". 

The  operator  then  adds  +20 dB 

to  the  INPUT/OUTPUT  RATIO  reading  and 
the  recorded  level. 

If  the  LEVEL  AT  switch  is  at  INPUT,  the 
actual  INPUT  voltage  of  BKF10  is 

10  times 

(20 

dB)  that  indicated  by  the  LEVEL  CON­

TROL  switch  (16). 

Example: 

MAX.  INPUT  switch  is  at  "30  V",  the 
"-dB"  and  " +20 dB"  lamps  are  lit,  and 

the  meter  reads  7  dB.  Then  the  true  IN­

PUT/OUTPUT  ratio  is:  -7  dB  +  20  dB 

+13  dB. 

OUT -OF-RANGE  LAMPS 

a.  If  the  INPUT  out-of-range  lamp  lights, 

check  MAX.  INPUT  switch  setting.  This 
will  occur  if  the  input  level  to  the 
BKF10  is  either  too  high  or  too  low. 

b.  If  the OUTPUT  out-of-range  lamp  lights, 

check  the  settings  of  the  level  control 
switch  (16),  VERNIER  (15)  and  LEVEL 
AT  switch  (17). 

c.  If  both  out-of-range  lamps  light,  this 

normally  indicates  that  either  there  is 
no  signal  path  through  the  device  un­
der  test  (LEVEL  AT  switch  at  INPUT 

only)  or the output level  from  the  device 

under  test  is  inadequate  for  making 
measurements. 

d.  If  either  of  the  out-of-range  lamps 

lights,  for  example,  during  a  sweep, 
check  the  characteristics  of  the  device 

under  test.  This  can  be  the  case  partic­
ularly  where  measurements  are  made 
at  a  constant  output  from  the  device 
under  test  (LEVEL  AT  switch  (17)  set 
to  INPUT),  and  where its characteristics 
prevent  the  B

'

KF10  from  delivering  an 

output  level  sufficient  to  maintain  the 
constant  output  required  from  the  de­
vice  under  test. 

HIGH-PASS  AND  LOW-PASS  FILTERS 

250  HZ 

Fig. 

3. 

Characteristics  of the  LP/HP-filters. 

1.  To  eliminate  the  effects  of  hum  and 

noise  below  250  Hz,  set  the  INPUT 
FILTER  (2)  switch  to  HP.  Note  that 
only  fundamental  frequencies  above 
250  Hz  are  relevant. 

2.  To  measure  distortion  due  to hum  and 

noise  below  250  Hz,  set  the  INPUT 

FILTER  switch  to  LP.  If  the  funda­

mental  frequency  is  above  500  Hz,  the 
"DISTORTION"  indicated  will  be  es­

sentially  due  to  hum.  When  measuring 

distortion  with  the  low-pass  filter,  only 
signals  with  a  fundamental  frequency 
below  approx.  80  Hz  are  relevant. 

3.  In  the  FLAT  setting  of  the  INPUT 

FILTER  switch,  both  filters  are  out 

of  action. 

USING  THE  DISTORTION  METER 

WITHOUT  THE  BUILT-IN  OSCILLATOR 

1.  Connect the  output,  for example,  of  an 

AF  generator  to  the  INPUT  of  the 
BKF10. 

2.  Check  the  input  level  (MAX.  INPUT 

switch  setting). 

3.  Read  off  the  distortion  factor  directly 

on  the  Distortion  Meter.  No  calibrat­
ing  or  balancing  is  necessary. 
The  INPUT  FREQUENCY  meter  will 
indicate  the  fundamental  frequency. 
The  INPUT/OUTPUT  Ri\TlO  meter  can 
be  used  to  indicate  the  signal  level 
in  dB  relative  to  the  setting  of  the 
LEVEL  CONTROL  switch  (16)  and  the 
VERNIER  (15),  provided  the  LEVEL  AT 
switch  is  set  to  OUTPUT. 
The  SIN  function  is  not  in  use  under 
these  conditions. 

If  it  is  desired  to 

.sweep 

the  AF  gener­

ator,  do  so  at  a  constant  rate  (per 

decade).  The  sweep  rate  should  not 
exceed  1  decade  per  20  seconds  for 
full  resolution. 

USING  THE  BUILT-IN  OSCILLATOR 

OUTPUT  WITHOUT  THE  DISTORTION 
METER 

The  BKF10  Oscillator  with  its  extremely 
low  distortion  «0.01%)  may  be  used  sep­
arately  as  an AF  signal  source. 

1.  Set  LEVEL  A  switch  (17)  to  OUTPUT. 

2.  Use  LEVEL  CONTROL  (16)  (and  VER­

NIER  (15),  if  necessary)  to  obtain  the 

output  level  required. 

3.  Set  oscillator  function  switoh 

(8) 

to 

either  SINGLE  SWEEP,  MAN.  or  EXT., 

as  required. 

4. 

To  monitor  the  frequency  (output  level 
above  10  mV  only),  use  a  BNC  T-con­

nector  and  connect  the  OUTPUT  to 
the  INPUT  with  a  short  length  of 
screened  cable.  See  Fig. 

4. 

SCREENED  CABLE 

F ig. 

4. 

OUTPUT 

Summary of Contents for BKF10

Page 1: ...the basic operat ing routine which is the same whatever the application so that the user will quick Iy become familiar with the BKF10 and its facilities A short description of the internal function o...

Page 2: ...3 V Sensitivity in 30 V 10 kfl setting 20 dB attenuator switched into input circuit 100 mV to 30 V 13 SIN pushbutton Depress button and read off SiN ratio directly on dB scale of Distortion Meter 14 L...

Page 3: ...by the LEVEL CON TROL switch 16 Example MAX INPUT switch is at 30 V the dB and 20 dB lamps are lit and the meter reads 7 dB Then the true IN PUT OUTPUT ratio is 7 dB 20 dB 13 dB OUT OF RANGE LAMPS a...

Page 4: ...PUT FREQUENCY dc analogue voltage of oscillator output frequen cy 3 V at 20 kHz to 0 V at 20 Hz 1 V decade f Expression Vdc log Volt 20 Hz Jack 23e INPUT FREQUENCY dc analogue voltage of INPUT FREQUEN...

Page 5: ...panel Pins 7 to 13 No connections Pin 14 3 V dc bias max 5 mAl for record ing see RECORDING below Equivalent to jack 21c X V RECORDER Recording distortion as a function of frequency 1 Connect the x a...

Page 6: ...front panel These fil ters allow the measurements of or the suppression of any hum and noise present in the input signal The frequency measuring section includes the frequency measuring circuit and t...

Page 7: ...20 Hz 20kHz SIN range Recorder output voltage HP LP filter Cult off frequency Attenuation Input output ratio meter Frequency range Measuring range Accuracy Recorder output voltage Sp V rms 50 dB appr...

Page 8: ...RADIOMETER COPENHAGEN _...

Reviews: