FRONT  PANEL  - CONTROLS  and  CONNECTORS 

"

I

'''' 

,..,. 

.-

.

1. 

Distortion meter - SiN meter.  I nd icates 
distortion  directly  in 

on  upper  scale 

and  SIN  ratio  in  dB  on  lower  scale. 

2.  HP-LP  filter  selector  switch.  Permits 

measurement  of,  or  suppression  of, 
hum  and  noise. 

3.  Line  power  ON/OFF  switch.  Red  lamp 

indicates  power  ON. 

4. 

Lamp  warns  when  indicated  ratio  is 
NEGATIVE  dB. 

5. 

Input/Output  Ratio  meter.  Indicates 

gain  or  loss  of  device  under  test  in 
dB. 

6.  Lamp  warns  when  MAX.  INPUT  switch 

is  set  to  30  V/10  kfl  setting.  Add  20 dB 
to  reading. 

7. 

Push button  for  start  of  built-in  single 
frequency  sweep  function. 

8.  Oscillator  function  selector  switch.  Se­

lects  SINGLE  SWEEP  function.  EX­

Ternal  function  or  MANual  freq.  set­
ting. 

OPERATING  INSTRUCTIONS 

The  identification  numbers  in  parentheses 
refer  to  the  circled  numbers  in  Figs.  1 
and  5. 

BEFORE  USE 

Before  connecting  the  BKF10  to  the  line 
supply,  ensure  the  following: 
Check  line  voltage  selector  (24)  and  line 
supply  fuse  (see  REAR  PANEL  CONNEC­

TORS  page  4).  To  change:  remove  plate, 
switch  to  alternative  line  voltage  and  re­

place  plate.  Line  fuse  values  are  printed 

below  fuse  holder. 

1. 

Plug  line  cord  into  receptacle  (26), 
connect  to  supply and depress POWER 
switch  (3)  to  switch  on.  Red  lamp  in­
dicates  power  ON. 

2.  If  the  instrument  is  to  be  used  on  a 

work  table,  pull  down  the  stand  fitted 
under  the  front  of  the  bottom  plate. 

It  will  be  found  easier  to  read  the 

meters  In  this  position. 

Fig. 

1. 

9. 

Input  Frequency  meter.  Indicates  the 
frequency  of  signal  present  at  the  in­

put  of  the  BKF10. 

10.  INPUT  connector  (BNC). 

11.  Lamp  warns  if signal  present  at  INPUT 

is  either  too  low  or  too  high.  Check 
setting  of  MAX.  INPUT  switch. 

12.  MAX.  INPUT  selector  switch.  Sensi­

tivity  in  3  V/1  Mfl  setting:  10  mV  to 
3  V.  Sensitivity  in  30  V/10  kfl  setting 
(20  dB  attenuator  switched  into  input 
circuit):  100  mV  to  30  V. 

13.  SIN  pushbutton.  Depress  button  and 

read  off  SiN  ratio  directly  on  dB  scale 
of  Distortion  Meter. 

14.  Lamp  warns  when  VERNIER  control  is 

turned  off  calibrated  (CAL)  position 

(fully  counter-clockwise). 

15.  VERNIER  control  for  level  control 

switch  settings.  Increases  OUTPUT  up 
to  10  dB  above  that  of  Level  Control 

switch  (16)  setting  (OUTPUT  deterio-

3.  Always  ensure  adequate  ventilation  for 

the  cooling  fins  (29)  on  the  rear  panel. 

4. 

The  GREEN 

jack  located  to  the 

right  of  the  line  fuse  is  used  to  ground 
the BKF10 if it  is  not  grounded through 
the  power  cord.  To  obtain  reliable 
measurements,  it  is  essential  that  the 
instrument  be  properly  grounded.  This 
jack  is  also  connected  to  the  centre 
pin  of  the  line  cord  receptacle  (26) 
and  to  pin  6  of  the  multi connector 

(28). 

Note: 

This  jack  should 

NOT 

be  used 

as  a  substitute  for  the  four  common 
signal  GROUND  jacks  (22). 

OU'''''  (,NO 

rates  above  1 V)  with  LEVEL  AT switch 

(17)  in  OUTPUT  position.  Increases  IN­
PUT  similarly  with  LEVEL  AT  switch 

(17)  in  INPUT  position. 

16.  LEVEL  CONTROL  switch.  Outer  scale 

refers  to  voltage  level  at  INPUT  con­
nector.  Inner  scale  refers  to  voltage 
level  at  OUTPUT  connector. 
EXT.  setting  permits  external  dc  ana­
logue  control  of  BKF10  output  or  input 

voltage. 

17.  Selector  switch  for  constant  INPUT 

level  to,  or  constant  OUTPUT  level 
from,  BKF10. 

18.  Manual  frequency  control  of  oscillator 

when  oscillator  function  switch  is  set 
to  MAN. 

19.  Lamp  warns  when  OUTPUT  from 

BKF10  is  too  low  or  too  high  (1  mV  to 
1  V).  Check  settings  of  MAX.  INPUT 
switch  and  LEVEL  CONTROL  switChl 
VERNIER. 

20.  OUTPUT  connector  (600 

Q) 

(BNC). 

BASIC  OPERATING  ROUTINE 

This  routine  uses  both  the Distortion  Meter 

and  the  built-in  oscillator  of  the  BKF10. 

1.  Connect  the  device  under  test  to  the 

BKF10,  using  screened  leads  as shown 

below. 

INPUT 

BKFIO 

OUTPUT 

DEVICE  UNDER 

,,:<J 

IN 

Fig. 2. 

Summary of Contents for BKF10

Page 1: ...the basic operat ing routine which is the same whatever the application so that the user will quick Iy become familiar with the BKF10 and its facilities A short description of the internal function o...

Page 2: ...3 V Sensitivity in 30 V 10 kfl setting 20 dB attenuator switched into input circuit 100 mV to 30 V 13 SIN pushbutton Depress button and read off SiN ratio directly on dB scale of Distortion Meter 14 L...

Page 3: ...by the LEVEL CON TROL switch 16 Example MAX INPUT switch is at 30 V the dB and 20 dB lamps are lit and the meter reads 7 dB Then the true IN PUT OUTPUT ratio is 7 dB 20 dB 13 dB OUT OF RANGE LAMPS a...

Page 4: ...PUT FREQUENCY dc analogue voltage of oscillator output frequen cy 3 V at 20 kHz to 0 V at 20 Hz 1 V decade f Expression Vdc log Volt 20 Hz Jack 23e INPUT FREQUENCY dc analogue voltage of INPUT FREQUEN...

Page 5: ...panel Pins 7 to 13 No connections Pin 14 3 V dc bias max 5 mAl for record ing see RECORDING below Equivalent to jack 21c X V RECORDER Recording distortion as a function of frequency 1 Connect the x a...

Page 6: ...front panel These fil ters allow the measurements of or the suppression of any hum and noise present in the input signal The frequency measuring section includes the frequency measuring circuit and t...

Page 7: ...20 Hz 20kHz SIN range Recorder output voltage HP LP filter Cult off frequency Attenuation Input output ratio meter Frequency range Measuring range Accuracy Recorder output voltage Sp V rms 50 dB appr...

Page 8: ...RADIOMETER COPENHAGEN _...

Reviews: