REAR PANEL - CONNECTORS
IV/OlCAOl
220\1
®
e
e-�
MY-et.,.
@
.......
Fig. 5.
21. Jacks for connection of external dc
analogue voltages for external control
of frequency and output level.
Ja:ck 21a.
OUTPUT FREQUENCY. 0 to +3 V dc
between jacks (21a) and (22) chang
es frequency at OUTPUT from 20
Hz to 20 kHz (1 V/decade), with os
oillator function switch (8) set to
EXT., input resistance >1 MO.
Expression: Vdc = log
[Volt)
20 Hz
Jack 21b.
AMPLITUDE. A dc analogue voltage
between jacks (21 b) and (22) chang
es level
as
given in I and
11
below
with LEVEL CONTROL switch (16)
set to EXT., input resistance> 1 MQ.
I. With LEVEL AT switch (17) set
to OUTPUT, the level at OUTPUT
changes from 1 V to 1 mV (1 V/de
cade) with analogue voltage of 0 to
-3 V dc.
Vouput
Expression: Vdc = log
1 V
[Volt)
11.
With LEVEL AT switch (17) set
to INPUT, level at INPUT changes
from 3 V to 10 mV (1 V/decade) with
an analogue voltage of + 0.48 V to
-2 V dc.
Expression: Vdc = log
[Volt)
Jack 21c.
3 V BIAS. dc bias voltage for re
cording, etc., output resistance <10,
max. load 5 mA.
lVlOlCAGI
22. Four common signal GROUND jacks.
23. Jacks for connection of dc analogue
output voltages to recorder, etc:
All output resistances <1 0, max.
loads 4 mA.
Jack 23d.
OUTPUT FREQUENCY, dc analogue
voltage of oscillator output frequen
cy: +3 V at 20 kHz to 0 V at 20 Hz,
1 V/decade.
f
Expression: Vdc = log
--
[Volt)
20 Hz
Jack 23e.
INPUT FREQUENCY, dc analogue
voltage of INPUT FREQUENCY me
ter: +3 V at 20 kHz to 0 V at 20 Hz,
1 V/decade.
f
Expression: Vdc = log -- [Volt)
20 Hz
Jack 23f.
DISTORTION
-
SIN
-I,
dc analogue
voltage of DISTORTION meter:
I. -4 V at 0.01% to -1 V at 10%,
1 V/decade.
11.
With SIN button depressed:
-4
V
at -80 dB to -1 V at
-
20 dB signal
to-noise ratio, 1 V/20 dB.
Expression: Vdc = log k [Volt)
k = distortion factor (100%-k=1)
Jack 23g.
INPUT/OUTPUT RATIO, dc analogue
voltage of INPUT/OUTPUT RATIO
meter: -2 V at -40 dB to +3.5 V at
70 dB, 1 V/20 dB.
Expression: Vdc = log
Vinput
[Volt)
Vouput
24. Line voltage selector.
25. Line supply fuse.
26. Line cord receptacle.
27. AF frequency output (BNC) from band
stop filter (harmonics only with funda
mental freq. rejected). Intended for
connection to oscilloscopes, etc.
28. Multiconnector. All dc analogue volt
ages, both for external control and
for recording, are also available here
for connection to accessories, e.g.,
the Radiometer REC61 S2 Servograph.
For details of the individual pin con
nections, see page 5.
29. Cooling fins. The BKF10 should be
placed so that adequate ventilation is
possible, particularly when mounted in
racks and cabinet systems.
30. Ground jack
-+-
This jack is used to
ground the instrument if it is not ground
ed via the power cord.