Warnings, Precautions Using the RS-2000 

 

Radiation Exposure  
 

As previously discussed, radiation exposure should be minimal unless the unit 
itself is compromised. Examples of unsafe conditions which could result in 
potentially dangerous radiation levels include; 

- Door interlock to the RS-2000 does not work 
- Shielding has been damaged 
- Evidence of machine tampering

 

 
 

 

 
 
 
 
 
 

 

Other Precautions 
 

- Another serious hazard from an x-ray device is electrical shock. The x-ray 
generator is a highly regulated DC power supply that applies a voltage of 
160 kilovolts. 
- No unauthorized personnel may defeat or override any safety features on 
the x-ray generator, collimator, or shielding, without permission of the 
manufacturer (Rad Source). The RS-2000 is to be serviced by trained 
personnel only. 
- Please keep unit dry. When cleaning do not allow cleaners or water to drip 
into panels or chamber. Use only a damp cloth with mild soaps for cleaning. 
- Do not use the top of the unit as a storage area. Do not place any heavy 
items or items containing liquids or materials that may harm the unit if they 
leaked or spilled on top or inside. 

The x-ray tube can be damaged if power is shut off 

immediately following an irradiation. After the 
irradiation is completed, wait 5 minutes before 
turning the unit off.

 

 

 

 

 
 

 

 

Turn power off

 

 

Summary of Contents for RS-2000

Page 1: ...egulatory Dose Limit General Hazards Associated with the X Ray Irradiator Safety Instructions Regarding the RS 2000 Irradiator Operating Instructions for the RS 2000 Irradiator 2 More complete and det...

Page 2: ...ct with the body So our next question is What is ionization and why is it a concern All matter is made up of atoms Atoms contain a small central nucleus containing protons which have electrically posi...

Page 3: ...called an ion The process of ionization is illustrated below This freed electron can potentially cause damage to living cells or DNA The damage can be repaired from small amounts of radiation however...

Page 4: ...generate x rays see picture below Basically what happens is that a metal filament is heated up to a very high temperature allowing electrons from the metal to essentially boil off Electrons the small...

Page 5: ...ill I receive from handling irradiated specimens and from using this irradiator As for handling irradiated specimens they do NOT become radioactive after an irradiation and you will not receive any ra...

Page 6: ...s No unauthorized personnel may defeat or override any safety features on the x ray generator collimator or shielding without permission of the manufacturer Rad Source The RS 2000 is to be serviced by...

Page 7: ...Indication Lights show when X rays are being produced Shielded and Interlocked X Ray Tube Chamber to prevent X ray production if panel should be opened Door Handle Double Safety Interlocked to prevent...

Page 8: ...ning module must be documented in order to get Xray access This training is different that the Radiation training orientation you need to complete both Radiation Safety for Radioactive Material Users...

Page 9: ...t Here are the different steps that may need to be taken Let me know which one fits your needs and I will send you the correct forms 1 If you currently have a mouse protocol and have mice housed in CC...

Page 10: ...gaged unit will not operate if not properly closed This can be done after you load your samples animals and set your time 2 Use the key to turn the power on off switch to on the key to power on the un...

Page 11: ...d that over heating can cause the unit to shut down I also have an electric fan in the room that you can set up to bring cooler air from the hall into the room 5 Main Menu The main menu enables the op...

Page 12: ...re two pre programmed warm up cycles labeled NORMAL and EXTENDED Unless you know that the unit hasn t been used for a couple of months select NORMALNORMAL 4 Press START at the bottom of the screen 5 T...

Page 13: ...re typically irradiated on Level 1 i e within the black RAD reflective chamber on the floor of the chamber Place no more than 5 mice in a CLEAN cage Remove food and water from the cage if you have exc...

Page 14: ...ormity is 98 The radiation dose rate is highly dependent upon the distance from the radiation source x ray tube Therefore for certain specimens e g test tubes flasks plates you should lay items horizo...

Page 15: ...S 29M 42S 22M 43S 16M 40S 11M 43S 7M 19S 3 500 35 31M 15S 30M 42S 34M 39S 26M 32S 19M 27S 13M 41S 8M 33S 4 000 40 35M 42S 35M 06S 39M 36S 30M 18S 22M 14S 15M 38S 9M 45S 6 000 60 53M 35S 52M 32S 59M 24...

Page 16: ...an irradiation time in for one programs A B C or D If you have already programmed in an irradiation time go to Step 3 After selecting PROGRAM SETUP the following screen AUTO PROGRAM SETUP will appear...

Page 17: ...afety Cabinet If the BSC is used spray and wipe down all surfaces inside the BSC Repeat with TEC Surf solution c Please dispose of all debris and waste in the waste basket provided d Empty the bleach...

Page 18: ...rrective actions for the RS 2000 Most of the time you will just need to hit clear and continue on If that doesn t work turn off the machine wait 20 seconds and turn it back on If neither of those opti...