room door while you are doing your run.  Please note that you have done this on 
the sheet in the room so that you are not double charged for your time.   
 
12. All users who sign up to use the X-ray on the online calendar have first priority 
and can ask a user to remove their sample if someone is in the room who has not 
signed up.  Please be courteous and allow a user to finish, most runs are short and 
a few minutes won’t hurt you.  Do note this on the sign-up sheet though, again I 
don’t want to over-charge if you have to wait. 
 
 
 
 

Faults & Possible Causes 

If during normal operation of the machine a fault condition is detected an ALARM 
SCREEN will be displayed.   Table 3 in the operating manual lists possible faults and 
the corrective actions for the RS-2000.  Most of the time you will just need to hit 
clear and continue on.  If that doesn’t work, turn off the machine, wait 20 seconds 
and turn it back on.  If neither of those options work, you will need to contact Deb 
at 

[email protected]

 
Keep in mind that some of the faults do not retain your time, so if you choose to 
leave the room and something happens your sample could be compromised. 
 
If you hit “stop” during a run, your time is lost and the machine will go back to the 
AUTO page.  Make sure to note your time prior to hitting stop. 
 
If you open the door during a run, your time is preserved and you can choose to 
initiate (Restart) your run or cancel it. 
 
 

Other Pertinent Information 

On the Rad Source website (

www.radsource.com

) information pertaining to the 

RS-2000 irradiator can be found. The website also provides contact information 
should you wish to contact a technical representative directly. 

 

Summary of Contents for RS-2000

Page 1: ...egulatory Dose Limit General Hazards Associated with the X Ray Irradiator Safety Instructions Regarding the RS 2000 Irradiator Operating Instructions for the RS 2000 Irradiator 2 More complete and det...

Page 2: ...ct with the body So our next question is What is ionization and why is it a concern All matter is made up of atoms Atoms contain a small central nucleus containing protons which have electrically posi...

Page 3: ...called an ion The process of ionization is illustrated below This freed electron can potentially cause damage to living cells or DNA The damage can be repaired from small amounts of radiation however...

Page 4: ...generate x rays see picture below Basically what happens is that a metal filament is heated up to a very high temperature allowing electrons from the metal to essentially boil off Electrons the small...

Page 5: ...ill I receive from handling irradiated specimens and from using this irradiator As for handling irradiated specimens they do NOT become radioactive after an irradiation and you will not receive any ra...

Page 6: ...s No unauthorized personnel may defeat or override any safety features on the x ray generator collimator or shielding without permission of the manufacturer Rad Source The RS 2000 is to be serviced by...

Page 7: ...Indication Lights show when X rays are being produced Shielded and Interlocked X Ray Tube Chamber to prevent X ray production if panel should be opened Door Handle Double Safety Interlocked to prevent...

Page 8: ...ning module must be documented in order to get Xray access This training is different that the Radiation training orientation you need to complete both Radiation Safety for Radioactive Material Users...

Page 9: ...t Here are the different steps that may need to be taken Let me know which one fits your needs and I will send you the correct forms 1 If you currently have a mouse protocol and have mice housed in CC...

Page 10: ...gaged unit will not operate if not properly closed This can be done after you load your samples animals and set your time 2 Use the key to turn the power on off switch to on the key to power on the un...

Page 11: ...d that over heating can cause the unit to shut down I also have an electric fan in the room that you can set up to bring cooler air from the hall into the room 5 Main Menu The main menu enables the op...

Page 12: ...re two pre programmed warm up cycles labeled NORMAL and EXTENDED Unless you know that the unit hasn t been used for a couple of months select NORMALNORMAL 4 Press START at the bottom of the screen 5 T...

Page 13: ...re typically irradiated on Level 1 i e within the black RAD reflective chamber on the floor of the chamber Place no more than 5 mice in a CLEAN cage Remove food and water from the cage if you have exc...

Page 14: ...ormity is 98 The radiation dose rate is highly dependent upon the distance from the radiation source x ray tube Therefore for certain specimens e g test tubes flasks plates you should lay items horizo...

Page 15: ...S 29M 42S 22M 43S 16M 40S 11M 43S 7M 19S 3 500 35 31M 15S 30M 42S 34M 39S 26M 32S 19M 27S 13M 41S 8M 33S 4 000 40 35M 42S 35M 06S 39M 36S 30M 18S 22M 14S 15M 38S 9M 45S 6 000 60 53M 35S 52M 32S 59M 24...

Page 16: ...an irradiation time in for one programs A B C or D If you have already programmed in an irradiation time go to Step 3 After selecting PROGRAM SETUP the following screen AUTO PROGRAM SETUP will appear...

Page 17: ...afety Cabinet If the BSC is used spray and wipe down all surfaces inside the BSC Repeat with TEC Surf solution c Please dispose of all debris and waste in the waste basket provided d Empty the bleach...

Page 18: ...rrective actions for the RS 2000 Most of the time you will just need to hit clear and continue on If that doesn t work turn off the machine wait 20 seconds and turn it back on If neither of those opti...