To remove the 1260-100/101 Carrier from the VXIbus chassis,
loosen the upper and lower retaining screws. Simultaneously push
the upper injector handle upward and pull the lower handle
downward until the carrier is disengaged from the chassis
backplane. Then slide the carrier out of the chassis.
MAKING
CONNECTIONS
TO PLUG-INS
(PIN-OUTS)
For information about connecting external equipment to the plug-
ins, refer to the individual manuals for the plug-ins. These include
information regarding mating connector part numbers, pinouts, and
electrical ratings.
POWER-ON
SELF-TEST
When the VXIbus chassis is switched on, the Option-01T
automatically runs a self-test. While this self-test is running, the
Fail LED on the front panel of the carrier illuminates. Figure 1-7
shows the location of the Fail indicator. If the carrier is not
equipped with an Option-01T, the Fail LED will not illuminate.
When the Option-01T completes self-test, the Fail LED will
extinguish. This occurs within five seconds after chassis power is
turned on.
The self-test includes the following:
1. PROM checksum
2. RAM write/read pattern verification
3. Microprocessor instruction test
4. Non-volatile RAM verification
5. Timer test
This self-test represents 70% fault detection.
CLOSING AND
OPENING
RELAYS
For general commands to close and open relays, refer to the
section
Closing a Relay
in
Chapter 1: Getting Started
of the
Option-01T Instruction Manual. Other commands are specific to
individual plug-ins. For detailed information on module-specific
commands, refer to the instruction manual for the appropriate plug-
in.