Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device’s warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers. Do not perform service and maintenance while the user is in the product.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
•
Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
Monthly
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Warewashing machine
This product can be washed at 60° using a mild
detergent for 10 minutes in a washing machine
designed for medical devices. Use the machine’s
drying function to dry the product.
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer’s
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• The frame height can be adjusted using the
spindle
• Threads for fasteners are in good condition
• Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
• Gas-spring for seat tilt is adjusted correctly
and is possible to lock the seat plane in
position
• Castors are moving freely and all wheel
locks/brakes are easily activated
• Nuts and bolts on the product are tightened
• Fittings and buckles for signs of worn out
parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
EN
Summary of Contents for Strong Base
Page 1: ...support R82 org 2020 03 rev 7 EN 12182 600 mm 23 120 kg 264 5 lb Strong Base User Guide ...
Page 3: ...3 support R82 org 120 kg 264 5 lb 120 kg 264 5 lb ...
Page 4: ...4 support R82 org 6 mm 13 mm 6 mm 13 mm 2 4 5 3 1 ...
Page 5: ...5 support R82 org 05 2 3 5 6 4 A B C A 30 41 cm B 35 46 cm C 40 51 cm 1 ...
Page 6: ...6 support R82 org 5 mm 5 mm 2 3 5 6 4 1 ...
Page 8: ...8 support R82 org 6 mm 13 mm 2 5 Nm 6 mm 13 mm 2 3 5 4 05 6 1 ...
Page 9: ...9 support R82 org 05 2 3 5 6 4 10 25 1 ...
Page 10: ...10 support R82 org 2 4 5 3 x 1 1 ...
Page 11: ...11 support R82 org x 2 3 2 4 5 1 ...
Page 12: ...12 support R82 org 4 mm 10 mm 2 5 Nm 2 3 5 6 4 1 ...
Page 13: ...13 support R82 org 9 10 7 8 1 x 2 x ...
Page 14: ...14 support R82 org 05 6 mm 2 5 Nm 6 mm 2 5 Nm 2 4 5 3 120 kg 1 ...
Page 15: ...15 support R82 org 17 mm 17 mm 3 2 4 1 ...
Page 16: ...16 support R82 org 4 4 4 4 A B A B XX XXXXXXXXXXXXXX XX XXXXXX XX XXXXXX ...
Page 76: ...support R82 org ...