Precautions to Take Before
Running the KOBRA
®
CELL
•
This device has only to be used with the supplied
power pack.
•
Always ensure there is a mobile phase flowing
through the KOBRA® CELL before switching on
the current source.
•
Do not switch on without mobile phase flowing
through the KOBRA® CELL or the membrane will
be damaged.
•
It is advised to run the mobile phase through the
KOBRA® CELL for 30 minutes prior to analysis.
•
Always turn off the KOBRA® CELL power pack
before turning off the HPLC pump.
•
Always use HPLC grade solvents (with minimal
benzene content) from a quality supplier and
maintain fresh supplies of potassium bromide.
•
Never flush 100 % solvent through the
KOBRA® CELL as this can damage the membrane.
4.6
Operating the KOBRA
®
CELL
The KOBRA® CELL power pack is turned on by
moving the switch to the ‘On’ position. To shut
down the device move the switch to the ‘Off’
position.
4.7
Error LED
•
The LED will light up if there is no KOBRA® CELL
attached to the current source.
•
The LED will also light up when the KOBRA® CELL
is dry and the flow of the current source is
inhibited.