COMPONENT
PLACEMENT
14
4
Place the laser printer or chart recorder on top of
the cart.
Note
: Follow printer manufacturer’s instructions to
set up the printer.
5
Plug the four component AC power cables into the
transformer. (See Figure 1 on page 4 for
appropriate plug placement.)
6
While holding the four AC cables, place the
transformer on the bottom of the cart.
7
Connect an AC cable to the transformer.
Connect the third standard length AC cable to the
computer tower. Connect the longest AC cable to
the pre-amp.
Route the two of the standard length AC cables to
the top of the cart for the monitor and chart
recorder.
Note: Make sure the preamp is 3 feet or farther
away from the isolation transformer. Electro-
magnetic interference could cause noisy ECG
signals.
8
Depending on the type of printer:
•
Connect the USB cable and the power cord to
the printer.
Or
•
Connect the parallel cable to the chart
recorder.
Summary of Contents for Q-Stress 4.6
Page 4: ...SCOPE 2...
Page 8: ...HARDWARE CONNECTIONS 6...
Page 10: ...SYSTEM INSTALLATION ON A STRESS CART 8...
Page 14: ...INSTALL THE WRITING SURFACE AND PAPER TRAY 12...