First commissioning step by step
To prevent incorrect parameter settings from causing defects in the Peltier element or controller, no
load must be connected before the first start-up.
Connect the supply voltage and the sensor to terminals 3 to 6 as described under Controller
connection. Please leave the remaining connections free.
Check your design and switch on the supply voltage.
Check the
ACTUAL TEMPERATURE
. value. If an unrealistic temperature is displayed, switch off the
supply voltage and please check your assembly and the sensor. The controller may only be operated
with a PT1000 sensor.
Switch off current flow
After the device is switched on, the MONITOR-menu appears on the display.
Make sure that the operation indicator
appears red as shown in the picture. The output is not
switched through. If the display lights up green, then tap this field
to switch off the output.
correct
incorrect
Parameter settings
Switch from the MONITOR to the parameter menu
To do this, tap
- You get into the password menu:
You will be asked to enter the password. By default, the password is
0000.