This parameter is currently valid only for Active Energy Total Import/Export (kWh), Reactive Energy Total (kvarh), Total Energy (kVAh)
Note: Device is reporting only if there was a change of 0.1 in EnergyNote: In the future will be possible to measure and report also Active Energy on PH1,
PH2 and PH3.
Size: 2 Byte, Default Value: 600
Setting
Description
0
reporting disabled
600 - 65536
Reporting enabled. Report is send with the time interval set by entered value. Seconds
Parameter 43: Other Values - Reporting on time interval
This parameter is valid only for Voltage (V of ph1, ph2, ph3), Current (A of ph1, ph2, ph3), Total Power Factor, Total Reactive Power (var)
Note: Device is reporting only if there was a change
Size: 2 Byte, Default Value: 600
Setting
Description
0
reporting disabled
600 - 65536
Reporting enabled. Report is send with the time interval set by entered value. Seconds
Parameter 100: Enable / Disable External IR relay (BICOM)
NOTE1: After parameter change, first exclude module (without setting parameters to default value) and then re include the module.NOTE 2: If you don‘t
have IR BICOM relay module mounted and you enable IR communication (parameter 100 is 1 or 2) there will be no valid IR relay state reported. It will be
reported IR COMMUNICATION ERROR and LED2 will BLINK.
Size: 1 Byte, Default Value: 0
Setting
Description
0
External IR relay disabled
1
External IR relay enabled and connected to all 3 Phases
2
External IR relay enabled and connected to a Phase 1
Parameter 101: Enable / Disable External relay (IKA)
After parameter change, first exclude module (without setting parameters to default value) and then re include the module.
Size: 1 Byte, Default Value: 0
Setting
Description
0
External relay disabled
1
External relay enabled and connected to Phase 2
Parameter 106: External IR relay (BICOM) power threshold settings – maximum power of all phases together
This parameter defines a threshold when External IR relay is being turned off. (If Parameter no. 100 is set to the value 1 or 2)
NOTE: Meter is capable of measuring max 3x65A!
Size: 2 Byte, Default Value: 0
Setting
Description
0
no function
10 - 60000
Watt
Parameter 107: External relay (IKA) power threshold settings – maximum power on phase L2
This parameter defines a threshold when External relay is being turned off (if the parameter no. 100 is set to the value 1 or 2).
NOTE: Meter is capable of measuring max 65A
Size: 2 Byte, Default Value: 0
Setting
Description
0
no function
10 - 20000
Watt
Parameter 112: Power threshold – Delay before power on
External IR relay/ External relay is turned off due to detected overload (as set by parameter 106&107) and remains off for a time, defined in this parameter.
After that time, the output turns on to check, if the overload is still present.