![Quantronix Cubiscan 110-XT Operation And Technical Manual Download Page 48](http://html1.mh-extra.com/html/quantronix/cubiscan-110-xt/cubiscan-110-xt_operation-and-technical-manual_3466065048.webp)
Cubiscan 110-XT
38
CHAPTER 5
CALIBRATION
This chapter provides instructions for calibrating the Cubiscan 110-XT. The
Cubiscan 110-XT is calibrated at the factory; however, some circumstances
in which recalibration may be required include the following:
•
Calibrate the Cubiscan 110-XT if you have problems cubing and
weighing after assembly and setup.
•
Calibrate the Cubiscan if it is subjected to any type of mechanical shock
or collision with a heavy object.
•
Calibrate the Cubiscan as part of a regular maintenance schedule. If the
Cubiscan is used heavily, scale calibration should be performed monthly
and sensor calibration yearly.
N
OTE
If an error message appears during calibration, power the Cubiscan off and back
on and start calibration over.
Before you begin
Before calibrating the Cubiscan 110-XT, remove all packages or other
material from the platform, and blow any dust off the sensor screens. Refer
to
for information on cleaning the sensors.
All controls and displays for the Cubiscan 110-XT are located on the
freestanding touchscreen. For information on the controls and indicators,
refer to
“Measuring and weighing using the touchscreen” on page 25
If you want to calibrate using Qbit, refer to the
Qbit User’s Guide
.
N
OTE
The calibration cube should be kept clean and undamaged—you will need it each
time you calibrate the Cubiscan 110-XT.