Calibration
Calibrating the ultrasound sensors
Cubiscan 110-XT
42
To calibrate the sensors using the touchscreen, proceed as follows.
1. At the home screen, tap
CALIBRATE
.
2. The calibration menu is displayed at the bottom of the screen. Select
the
Sensors
option if it is not already selected.
3. Make sure there is nothing on the Cubiscan 110-XT platform, and tap
[Next]
to continue.
Figure 36
Home screen
Figure 37
Sensors calibration menu