Recommended settings
Sample name
and type
Bead matrix (lysing
matrix)
Setting
(speed – number of cycles x time – pause)
Application
Microbial
Bacteria
Garnet 0.15 mm
2000 – 1 x 5 min – 0 s
DNA isolation
Bacteria
Glass 0.1 mm
2000 – 1 x 5 min – 0 s
DNA isolation
Animal
Muscle
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Brain
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Liver
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Kidney
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Lung
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Heart
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Spleen
Ceramic 2.8 mm
3500 – 2 x 45 s – 30 s
RNA isolation
Plant
Leaves
Ceramic 2.8 mm
4200 – 2 x 45 s – 30 s
RNA isolation
Leaves
Metal 2.38 mm
2,000 – 1 x 3 min
DNA isolation
Seeds
Metal 2.38 mm
2,900 – 1 x 3 min
DNA isolation
Stems
Ceramic 2.8 mm
4,200 – 2 x 45 s – 30 s
RNA isolation
Stems
Metal 2.38 mm
2,200 – 1 x 3 min
DNA isolation
Roots
Ceramic 2.8 mm
4200 – 2 x 45 s – 30 s
RNA isolation
Root
Metal 2.38 mm
2500 – 1 x 3 min
DNA isolation
Flowers
Ceramic 2.8 mm
4200 – 2 x 45 s – 30 s
RNA isolation
Soil
Soil
Garnet 0.7 mm
2000 – 1 x 5 min
DNA isolation
Soil
Glass 0.1 mm
Clay: 4000 – 1 x 45 s – 0 s
Loose, high humic: 2500 – 1 x 45 s – 0 s
DNA isolation