PowerLyzer 24 Homogenizer User Manual 09/2018
From this screen, the user can select the display language by pressing and holding
to
display the language. The user can then choose from English, Italian, French, German or
Spanish by pressing the
and
arrow keys.
After 5 s, a menu displays the programming mode. The programming mode displays the last
program run before the power was turned off. An example is shown below.
PROG>01
20°C
S=4000
C=02
T=0:10
D=0:10
READY TO START
1
Program number
5
Chamber temperature
2
“S=” denotes speed of the motor ranging
from 500 to 5000 rpm, in increments of
100 rpm.
6
“C=” denotes the number of times the
cycle (speed x time) is repeated. The
variable can range from 01 to 10, in
increments of 1.
3
“T=” denotes the time interval in min:s.
The variable can range from 0:01 to
9:59, in increments of 1 s.
7
“D=” denotes the time interval in seconds
of dwell/pause between each cycle. The
variable can range from “Dmin” (Dwell
minimum* expressed as min:s) to 9:59,
in increments of 1 s.
4
System status
* Dwell minimum varies depending upon the speed and length of cycle.
1
2
3
5
6
7
4