7
Calibration
7.1
PTSA Calibration (Two-Point with Zero)
1. Rinse the main module sample cell three times with DI water. Fill the sample cell with DI water.
*
NOTE
*
In emergency, “non-PTSA” water, such as city water, may be used, but re-
calibrate using DI water for the zero step as soon as it is available.
2. Power on the SP-710B by pressing
. Allow 5–10 seconds for the SP-710B to stabilize.
3. The unit is actively reading and displaying both PTSA and Conductivity. The values will be very low if
DI water is used; conductivity value is not critical but PTSA value should be near zero. A low non-zero
value (e.g. 0.2 or 0.4, etc.) is not problematic.
4. Press
Calib
(
) as needed to highlight
PTSA
in the selection menu.
5. Press
to launch the
PTSA CALIBRATION
screen (Figure 10).
6. Press
Zero
(
) to start the zero (blank) calibration.
7. If the calibration succeeds, a checkmark (
) and instructions for the slope calibration will appear
(Figure 11).
8. Rinse the main module sample cell three times with the desired PTSA standard. Fill the sample cell
with the desired PTSA standard.
9. Press
Cycle
(
) to cycle between the PTSA standards 100, 200, and 300 ppb (it repeats). Ensure
the value selected matches the desired PSTA standard in the sample cell.
10. Press
Slope
(
) to start the slope calibration.
11. If the calibration succeeds, a checkmark (
), a ”
Calibration Success
” message will appear (Figure
12). Otherwise, a warning message is displayed.
12. Calibration is now complete. Press
Exit
(
) to return to measurement mode.
*
NOTE
*
If
Exit
is pressed before the second checkmark appears, the calibration will
not be completed and must be re-done.
Figure 10.
Figure 11.
Figure 12.
SP-710B User Manual
| +1 (866) 203-8397
14
Summary of Contents for SP-710B
Page 1: ......