8.1
Main Module Sample Cell Cleanliness Check
The SP-710B is designed to provide reliable and accurate measurement on PTSA and Fluorescein. Heavy
fouling will prevent the light from reaching the sensor, resulting in inaccurate readings. It is suggested that
the SP-710B be checked for fouling and cleaned on a monthly basis. Heavily contaminated waters may
require more frequent cleanings. Cleaner water sources with less contamination may not require cleaning
for several months. The SP-710B is designed to carry out a Cleanliness Check as described below:
1. Power on the SP-710B by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Diagnosis
(
) to launch the
SYSTEM DIAGNOSIS
screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from
Starting BTLE...
to
BTLE Started
5. Press
Cleanliness
(
). An instruction prompt appears to ask the user to put DI water into the main
module sample cell (Figure 33).
6. Pour DI water into the main module sample cell.
7. Press
Confirm
(
,
, or
). The instruction prompt will disappear and the SP-710B displays
a countdown toward the bottom of the display.
8. Once the Cleanliness Check is completed a
Clean
message (Figure 34) or
Sample cell fouled
mes-
sage (Figure 35) will appear towards the bottom of the display.
9. Cleanliness check is now complete. Press
Exit
(
) to return to measurement mode.
Figure 33.
Figure 34.
Figure 35.
8.2
Bluetooth Connection to Devices
The SP-710B uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone
via the
uPyxis®
Mobile App, to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the
uPyxis®
Desktop App, or to nearby Pyxis inline probe with the Pyxis Inline Bluetooth Adapter (P/N: MA-WB)
connected between the Pyxis inline probe and controller. To allow the SP-710B to connect via Bluetooth
with other devices, follow the steps below: