(a) Press
Continue
(
or
) to proceed to a four-point calibration,
or
(b) Press
Exit
(
) to end the calibration process at a three-point calibration and return to mea-
surement mode.
7.4.3
Four-Point Calibration
15. Rinse the main module sample cell three times with 500 ppb Fluorescein standard. Fill the sample
cell with the standard.
16. Press
Slope
(
or
) to start 500 ppb slope calibration.
17. If the calibration succeeds, a second checkmark (
) ”
Calibration Success
” message will appear
(Figure 25). Otherwise, a warning message is displayed.
18. Calibration is now complete. Press
Exit
(
) to return to measurement mode.