1. Power on the SP-710 by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Diagnosis
(
) to launch the
SYSTEM DIAGNOSIS
screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from
Starting BTLE...
to
BTLE Started
(Figure 50).
5. Choose to connect via one of two options:
(a) The
uPyxis®
Mobile App (see the
section), or
(b) The
uPyxis®
Desktop App (see the
section).
8.2.1
Calibrate an ST-500 Series Probe with the SP-710 via Bluetooth
The SP-710 can be used to verify the result of an inline Pyxis ST-500 Series probe by measuring the sample
water taken from the inline probe sample line. The SP-710 can then be used to calibrate the inline probes
over the Bluetooth connection. To calibrate an inline probe, follow the steps below:
1. Power on the SP-710 by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Comm
(
) to launch the
COMMUNICATION
screen (Figure 54).
4. Press
Scan
(
) to begin scanning for Bluetooth devices.
5. Discoverable devices will begin to populate on the display with their name and MAC-Address (Figure
6. If more than one device appears in the
Device list
, press
»
(
) to cycle through the devices.
7. If no devices or the incorrect device appear in the
Device list
, press
Scan
(
) to re-scan for discov-
erable devices.
8. Press
Connect
(
) to begin pairing to the selected probe.
9. When the connection is established, the SP-710 displays the latest PTSA measurement from the con-
nected probe (Figure 56).
10. Fill the main module sample cell with the same sample water that the probe is measuring.
11. Press
Read
(
) to see PTSA measurement from the SP-710along with the probe measurement
(Figure 57).
12. Press
Calib
(
) to begin probe PTSA calibration.
13. The SP-710 will take the probe PTSA measurement three times to verify the calibration (Figure 58).
*
NOTE
*
It takes about one minute for the probe to approach the calibrated reading
and the three verifying readings may not be exactly the same as the value measured
by the SP-710. Press
Read
(
) again to take more readings from the probe, if
necessary.