Main Module Sample Cell Cleanliness Check
The SP-710 is designed to provide reliable and accurate measurement on PTSA and TMB Free and Total
Chlorine. Heavy fouling will prevent the light from reaching the sensor, resulting in inaccurate readings. It
is suggested that the SP-710 be checked for fouling and cleaned on a monthly basis. Heavily contaminated
waters may require more frequent cleanings. Cleaner water sources with less contamination may not re-
quire cleaning for several months. The SP-710 is designed to carry out a Cleanliness Check as described
below:
1. Power on the SP-710 by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Diagnosis
(
) to launch the
SYSTEM DIAGNOSIS
screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from
Starting BTLE...
to
BTLE Started
5. Press
Cleanliness
(
). An instruction prompt appears to ask the user to put DI water into the main
module sample cell (Figure 51).
6. Pour DI water into the main module sample cell.
7. Press
Confirm
(
,
, or
). The instruction prompt will disappear and the SP-710 displays a
countdown toward the bottom of the display.
8. Once the Cleanliness Check is completed a
Clean
message (Figure 52) or
Sample cell fouled
mes-
sage (Figure 53) will appear towards the bottom of the display.
9. Cleanliness check is now complete. Press
Exit
(
) to return to measurement mode.
Figure 51.
Figure 52.
Figure 53.
8.2
Bluetooth Connection to Devices
The SP-710 uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone
via the
uPyxis®
Mobile App, to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the
uPyxis®
Desktop App, or to nearby Pyxis inline probe with the Pyxis Inline Bluetooth Adapter (P/N: MA-
WB) connected between the Pyxis inline probe and controller. To allow the SP-710 to connect via Bluetooth
with other devices, follow the steps below:
SP-710 User Manual
| +1 (866) 203-8397
29
Summary of Contents for 50352
Page 1: ......