12. Press
Calib
(
) to start slope calibration.
13. The message “
Slope calibration success
” will appear on the bottom of the display (Figure 43).
14. Calibration is now complete. Long press
Calib
(
) to return to the
CHLORINE MEASUREMENT
screen.
Figure 38.
Figure 39.
Figure 40.
Figure 41.
Figure 42.
Figure 43.
7.4.5
Restore Default Chlorine Calibration Parameters
Pressing the
Default
(
) during any of the chlorine calibration procedures will copy the default calibration
intercept and slope to the working intercept and slope, respectively. This button action is to restore the
working calibration parameters to the original, factory-loaded calibration parameters for both free and total
chlorine.