28
sensor readings are stable.
Note: If using retractable needle-type sensors (e.g.
OXR50
,
OXR230
,
OXR430
), it is important that the sensor tip is extended!
Then press SET 0% and the actual oxygen sensor
reading is taken for the 0% calibration. After the
completion of the 0% calibration, the system
returns to the Calibration Menu, showing now the
date of the 0% calibration.
Note: SET 0% is only active if the oxygen readings are within the
expected range for the connected sensor type. If SET 0% is
replaced by "out of range", check or replace the calibration
standard and repeat the calibration.
6.4
Options Menu
In the Options Menu several device and sensor
parameters can be adjusted. Furthermore, device
information can be retrieved and a reset of the
device can be performed.
6.4.1
Adjust Time and Date
Please note that it is recommended to Adjust the local Time and
Date before the calibration of the connected sensor (see chapter
6.3) and before logging is started (see chapter 7.1). Adjust them by
moving the cursor with the BACK and OK buttons, and by
adjusting each item with the UP and DOWN buttons.