6.2.2
Medium
Adjust here the medium Gas or Water of your sample.
6.2.3
Oxygen Units
Select here your preferred oxygen unit. Note, that the available
oxygen units depend on the chosen medium.
6.2.4
Temperature
Here the temperature of the sample has to be
adjusted. This is important in order to assure a
correct temperature compensation of the oxygen
measurement.
Select Sensor for enabling automatic temperature compensation
based on temperature measurements of the connected PT100
temperature sensor (chapter 3.1.3).
In order to ensure proper automatic temperature compensation,
position the temperature sensor in the sample, so that it measures
the same temperature as given around the oxygen sensor tip.