![pyroscience FireStingGO2 Manual Download Page 24](http://html1.mh-extra.com/html/pyroscience/firestinggo2/firestinggo2_manual_1632880024.webp)
24
the 0% calibration is taken from the factory calibration encoded in
the Sensor Code.
It is recommended to perform the calibration at conditions close to
the environmental conditions during the measurements, especially
regarding the temperature.
6.3.1
Sensor Code
For security the Calibration menu repeats the Sensor Code entry,
which is in fact identical to the Sensor Code entry given in the
Settings menu (chapter 6.2.1).
Before starting the calibration, double check that the correct
Sensor Code has been entered, as written on the label attached to
the sensor cable or on the bag/stickers of the sensor.
6.3.2
Air Calibration
Select Calibrate Air for starting an air calibration.
Then
choose
the
type
of
temperature
compensation during the calibration:
a) For automatic temperature compensation of
the
Calibration
Temperature
with
the
temperature
sensor
connected
to
the
FireStingGO
2
meter, select Sensor and confirm
with Ok.
b) A Fixed Calibration Temperature can be
entered with the UP and DOWN buttons after the
temperature of the calibration standard was
measured e.g. with a third-party temperature
meter.
NOTE: Ensure constant calibration conditions! If the calibration is
performed with a Fixed temperature, the temperature in the
calibration standard must be measured and kept constant!