![Pyramid Technical Consultants F3200E User Manual Download Page 39](http://html1.mh-extra.com/html/pyramid-technical-consultants/f3200e/f3200e_user-manual_1632602039.webp)
PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 39 of 107
Figure 20. Strip chart display
When the data has filled the display, a slider appears underneath. You can use this to scroll back
through the data that has been buffered by the PTC DiagnosticG2 program (up to 100,000
readings). If you click in the graphic area near a trace, a cursor is placed with readout of the
channel number, the timestamp and the current at that point.
11.4.2 Scope display
A data block that has been buffered and averaged in the F3200E is displayed as a whole, in a
similar manner to a digital oscilloscope transient capture. The next available data block replaces
the previous one as a whole. This display is most suited to the display of buffered data,
especially the sweep mode.
In buffered mode the number of readings per refresh of the display is given by the burst size. If
no burst is defined, the screen is refreshed for each block of data sent up by the F3200E. In
sweep mode, the screen is refreshed with each averaged block of data.